A NEW VALIDATED RP-HPLC METHOD FOR THE ESTIMATION OF MYCOPHENOLATE MOFETIL IN PURE AND TABLET DOSAGE FORM

A. LAKSHMANA RAO, P. VIJAY SRINIVAS J.V.L.N.S. RAO

Abstract


A simple, rapid, sensitive and precise high performance liquid chromatographic (HPLC) method has been developed for the estimation of mycophenolate mofetil in bulk and tablet dosage form. In this method RP-C18 column (250mmx4.6mm I.D., 5μm particle size) with mobile phase consisting of acetonitrile and 0.03M phosphate buffer in the ratio of 60:40 v/v in isocratic mode was used. The detection wavelength is 254 nm and the flow rate is 0.8 ml/min. In the range of 20-120 μg/ml, the linearity of mycophenolate mofetil shows a correlation coefficient of 0.9995. The proposed method was validated by determining sensitivity, accuracy, precision and system suitability parameters. The mobile phase selected for the method is simple, fast, accurate and precise and hence can be applied for routine quality control analysis of mycophenolate mofetil in bulk and its tablet dosage form.

Refbacks

  • There are currently no refbacks.




 

 

Creative Commons License

This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 3.0 Unported License.

 

Listed as Green Publisher

Powered by Informatics India Ltd