A NEW VALIDATED RP- HPLC METHOD FOR DETERMINATION OF PENCICLOVIR IN HUMAN PLASMA

VENKATA KUMAR, ANANTAHAKUMAR, SESHAGIRI RAO

Abstract


A rapid, selective and sensitive high performance liquid chromatographic method for the estimation of Penciclovir in human plasma has been developed. Chromatography was carried out on a Hypersil BDS C18 column using a mixture of 20mM Potassium dihydrogen phosphate buffer (pH 3.5 + 0.05) and methanol (95:5 v/v) as the mobile phase. The eluents were monitored for the drug by UV detection at 254 nm. Acyclovir was used as an internal standard for this study. The retention times for Penciclovir and Acyclovir were found to be 6.2 and 9.2 min respectively. The method was found to be linear in the concentration range of 50.1 ng/mL to 7000.9 ng/mL. The method was validated as per FDA guidelines and was found to be suitable for bioequivalence and pharmacokinetic studies.

Full Text:

PDF

Refbacks

  • There are currently no refbacks.




 

 

Creative Commons License

This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 3.0 Unported License.

 

Listed as Green Publisher

Powered by Informatics India Ltd